Mindray TRIO Operations Instructions Page 166

  • Download
  • Add to my manuals
  • Print
  • Page
    / 170
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 165
Trio™ Operating Instructions 0070-10-0666-01 6 - 27
Appendix Physical Specifications
Isolation: Risk (Leakage) Currents:
Meets the requirements of IEC 60601-1/
EN 60601-1
Enclosure Risk Current:
Normal operating conditions <= 100 μA
Single fault condition <= 300 μA for 120 VAC
Single fault condition <= 500 μA for 230 VAC
Patient Source Current:
Normal operating conditions <= 10 μA
Single fault condition <= 50 μA
Patient Sink Current:
Normal operating conditions not applicable
Single fault condition <= 50 μA
Dielectric Withstand Per IEC 60601-1/EN 60601-1