12-5
12.4 NIBP Setup Menu
Selecting the NIBP label in the parameter area opens the following menu.
Figure 12-2 NIBP Setup Menu
You can perform the following settings in this menu.
ALM NIBP alarm on/off status
ON: When a NIBP alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a NIBP alarm occurs, the monitor neither
gives alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the NIBP label.
ALM LEV Alarm level
Options: HIGH and MED.
ALM REC Alarm recording
ON: When an NIBP alarm occurs, the monitor enables the
recording;
OFF: When a NIBP alarm occurs, the monitor does not
enable the recording.