15-6
Alarm limits:
Parameter Upper Limit Lower Limit Step
CO2 (lower limit+2) ~100 0~(upper limit-2) 1
INS 0~100 / 1
AwRR (lower limit+2) ~150 0~(upper limit-2) 1
Default alarm limits:
Parameter Patient type Upper limit Lower limit
Adult 50 15
Pediatric 50 20
CO2
Neonate 45 30
Adult 4 /
Pediatric 4 /
INS
Neonate 4 /
Adult 30 8
Pediatric 30 8
AwRR
Neonate 100 30
NOTE
z The apnea alarm cannot be disabled.
z When the CO2 monitoring is not required, it is recommended the water trap not
be connected and the work mode be set to STANDBY.